Software safety classification
 ...

Software installed in medical devices is assessed for health and safety issues according to international standards.

Safety classes

Software classification is based on the potential for hazard(s) that could cause injury to the user or patient.[1]

Per IEC 62304|IEC 62304:2006 + A1:2015, the software can be divided into three separate classes:

  • The SOFTWARE SYSTEM is software safety class A if:
    • the SOFTWARE SYSTEM cannot contribute to a HAZARDOUS SITUATION; or
    • the SOFTWARE SYSTEM can contribute to a HAZARDOUS SITUATION which does not result in unacceptable RISK after consideration of RISK CONTROL measures external to the SOFTWARE SYSTEM.
  • The SOFTWARE SYSTEM is software safety class B if:
    • the SOFTWARE SYSTEM can contribute to a HAZARDOUS SITUATION which results in unacceptable RISK after consideration of RISK CONTROL measures external to the SOFTWARE SYSTEM and the resulting possible HARM is non-SERIOUS INJURY.
  • The SOFTWARE SYSTEM is software safety class C if:
    • the SOFTWARE SYSTEM can contribute to a HAZARDOUS SITUATION which results in unacceptable RISK after consideration of RISK CONTROL measures external to the SOFTWARE SYSTEM and the resulting possible HARM is death or SERIOUS INJURY“

Serious injury

edit

For the purpose of this classification, serious injury is defined as injury or illness that directly or indirectly is life-threatening; results in permanent impairment of a body function or permanent damage to a body structure; or necessitates medical or surgical intervention to prevent permanent impairment of a body function or permanent damage to a body structure.

References

edit
  1. ^ "Developing Medical Device Software to IEC 62304 | MDDI Medical Device and Diagnostic Industry News Products and Suppliers". www.mddionline.com. June 2010. Retrieved 2017-06-15.
Zdroj:https://en.wikipedia.org?pojem=Software_safety_classification
Text je dostupný za podmienok Creative Commons Attribution/Share-Alike License 3.0 Unported; prípadne za ďalších podmienok. Podrobnejšie informácie nájdete na stránke Podmienky použitia.






Text je dostupný za podmienok Creative Commons Attribution/Share-Alike License 3.0 Unported; prípadne za ďalších podmienok.
Podrobnejšie informácie nájdete na stránke Podmienky použitia.

Your browser doesn’t support the object tag.

www.astronomia.sk | www.biologia.sk | www.botanika.sk | www.dejiny.sk | www.economy.sk | www.elektrotechnika.sk | www.estetika.sk | www.farmakologia.sk | www.filozofia.sk | Fyzika | www.futurologia.sk | www.genetika.sk | www.chemia.sk | www.lingvistika.sk | www.politologia.sk | www.psychologia.sk | www.sexuologia.sk | www.sociologia.sk | www.veda.sk I www.zoologia.sk